ATTACHMENTS (JIGS)

Caution when choosing the right attachment
○Testing load has to be lower than the maximum load of the attachment.
○The weight of the upper attachment has to be lower than the 10% of the maximum load of the loadcell.
If the maximum load of the loadcell is very small, it is recommended to adjust the testing machine with the attachment attached being placed.
○Make sure the size of the connecting rod is correct.
Chart below shows the size of the screw for upper/lower attachment.
○If you plan to test special materials, it is highly recommended to perform the sample testing to choose the right attachment.


Installation size list
1:The installation of the upper jig
The upper jig is installed to a loadcell.
The screw size is as follows.
The machine model to use a loadcell equal to or less than 500N is for a male screw.
The jig side gets the female screw.
The machine model to use loadcell more than 1kN is for a female screw.
The jig side gets the male screw.
Load Under 500N 1kN〜10kN 20kN 50kN 100kN
Type SV-55C
SVZ-50NB
SVZ-200NB-50
SDT-203NB-50
SVZ-200NB-200
SDT-203NB-200
SDT-503NB
SDWS-0213
SDWS-0513
SDWS-1013
SDW-1003
SDWS-2013
SDW-2003
SDW-5003 SDW-9103
Screw size M6P1 M12P1.75 M16P1.5 M24P2 M32P2
※When a jig of the small capacity is installed with the additional load range, a connection screw is necessary.

2:The installation of the lower jig
The surface plate with T-Groove to fix lower jig is situated on the base of testing machine and the push-pull stand.
Two T-throw nuts are to be inserted into the T-Groove and fix a jig onto them with two bolts.
The dimension of the T-Groove and the size of the bolt are as follows.
Type SV-55C
SVZ-50NB
SVZ-200NB
SDT-203NB
SDT-502NB
SDWS-0213
SDWS-0513
SDWS-1013
SDWS-2013
SDW-1003
SDW-2003
SDW-5003 SDW-9103
T-Nut M8P1.25 M8P1.25 M8P1.25 M12P1.75 M16P2 M20P2.5
T Groove