| Selection of an attachment is very important, and it will bring out the full-capability of the testing machine. Almost every IMADA testing machine can measure the both "tensile" and "compression". Therefore the test of "tension", "compression", "bending", "stripping", and so on is possible by selecting the correct attachment for the testing purpose. We have more special attachement besides the items listed in this web site. We design the attachment which meets your requirement. Please feel free to contact us and let us know your requirement. |
| ¡Caution when choosing the right attachment | |
| | Testing load has to be lower than the maximum load of the attachment. |
| | The weight of the upper attachment has to be lower than the 10% of the maximum load of the load cell.If the maximum load of the load-cell is very small, it is recommended to adjust the testing machine with the attachment attached. |
| | Make sure the size of the connecting rod is correct. Chart below show the size of the screw for upper/lower attachment. |
| | If you plan to test unique and special materials, it is highly recommended to perform the sample testing to choose the right attachement. |
| ¡SELECT FROM THE LIST BELOW |
| Film Grip | Flat Chuck | Drill Chuck(Jacob's Chuck) |
| Wire Winding Grip | Wedge Grip | Tape Grip |
| Pantagraph Grip |
| Bend Test Stand | Flat Head Compression Attachment | Miscellaneous Attachement |
| Vise Grip | 90 degree peel table | |
| Chip Shear Test Attachment/Wire Terminal Grip/Paper Chuck | ||
| X-Y Table, Y-Z Table | ||
| Special Attachment Exsamples | ||
| ¡Regarding the mounting of the attachment |
| It is necessary to mount the attachment to our Testing Machine. And to bring out the full-capability of the testing machine, it is necessary to choose the right attachment. Therefore, attachment needs to be selected from our list or manufactured by customers. Below is the chart that shows a mating relations of our attachment. Please refer to the following charts when you mount the attachment or when you manufacture the attachment by yourself. Mounting Upper Attachment Upper attachment is to be mounted by screwing it into the load cell. Size of the screws are as follows. |
| Max. Load | Less than 500N | 1kN to 10kN | 20kN | 30kN to 50kN | 100kN |
| Model | Push-Pull Scale Digital Force Gauge SV-55C SV-52NA SV-52CA SDT-52NA SDT-52CA |
SV-100NA SV-100CA SV-200NA SV-200CA SV-201NA SV-201CA SV-301NA SV-301CA SDT-201NA SDT-201CA SVF-500NA SVF-500CA SDWS(up to 10kN) SDW-1002 SDW-1012 |
SDWS-2002 SDWS-2012 SDW-2002 SDW-2012 |
SDW-3002 SDW-5002 |
SDW-9902 |
| Size of the Screw | M6P1 | M12P1.75 | M16P1.5 | M24P2 | M32P2 |
*Size of screw for the Push-Pull Scale and Digital Force Gauge with the maximum load range from 1kN to 3kN is M10P1.5. |
| Mounting Lower Attachment Surface table is pre-installed for our Testing Machines and Push-Pull Stands. T-shaped channel is designed to mount the lower attachment by using 2 pairs of T-nuts and bolts. Shape of the T-shaped channel and the size of the bolts are as follows; |
| Model | SV-5 SV-55C SV-52NA SV-52CA SDT-52NA SDT-52CA |
SV-100NAA SV-100CA SV-200NAA SV-200CA SV-201NAA SV-201CA SV-301NAA SV-301CA SDT-201NAASDT-201CA SVF-500NAASVF-500CA |
SDWS-0202 SDWS-0212 SDWS-0502 SDWS-0512 |
SDWS-1002 SDWS-1012 SDW-1002 SDW-1012 SDW-2002 SDW-2012 |
SDW-3002 SDW-5002 |
SDW-9902 |
| Internal thread size of T-nuts | M8P1.25 | M8P1.25 |
M8P1.25 | M12P1.75 | M16P2 | M20P2.5 |
| Size of T-shape Channel | ![]() |
![]() |
![]() |
![]() |
![]() |
![]() |